Influence of the Site of Port Insertion on the Flow Velocity of 5-FU Continuous Intravenous Injection by Using a Compression-Type Portable Continuous Infuser for Colorectal Cancer Patients
Summary
For the continuous intravenous injection of fluorouracil(5-FU)over 46 hours in a chemotherapeutic regimen for colorectal cancer, a compression-type portable continuous infuser is used. However, there is an error in the flow velocity. Errors in the dose or time of administration completion may influence the efficacy and safety of 5-FU. Therefore, it is necessary to reduce the error in the flow velocity. The type of diluent, air temperature, and concentration of 5-FU are reportedly factors influencing the flow velocity. However, no study has examined the influence of the site of port insertion. In this study, we investigated factors associated with the flow velocity of 5-FU continuously and intravenously injected by using an infuser in patients with colorectal cancer, among whom the site of port insertion differed. The results showed that the site of port insertion influenced the flow velocity during the continuous intravenous injection of 5-FU using an infuser. It is important to reduce the error in the flow velocity by considering the site of port insertion in addition to the concentration of 5-FU and the air temperature.
要旨
大腸がんに対する化学療法であるフルオロウラシル(5-FU)の46時間持続静注のデバイスとして,加圧式携帯型持続注入器(以下,インフューザー)が用いられるが,流速に誤差が生じることが知られている。投与量または投与終了時間の誤差は5-FUの有効性や安全性に影響する可能性があるため,流速の誤差を極力低減することが求められる。流速への影響因子として希釈液の種類,気温および5-FU濃度などが報告されているが,ポート留置部位の影響について検討した報告はない。そこで,本研究ではポート留置部位が異なる大腸がん患者を対象とし,インフューザーによる5-FU持続静注の流速に影響する因子を検討した。その結果,ポート留置部位はインフューザーによる5-FU持続静注の流速に影響する因子であることが明らかとなった。5-FU濃度や気温以外にポート留置部位も考慮し,流速の誤差を低減することが重要である。
目次
Summary
For the continuous intravenous injection of fluorouracil(5-FU)over 46 hours in a chemotherapeutic regimen for colorectal cancer, a compression-type portable continuous infuser is used. However, there is an error in the flow velocity. Errors in the dose or time of administration completion may influence the efficacy and safety of 5-FU. Therefore, it is necessary to reduce the error in the flow velocity. The type of diluent, air temperature, and concentration of 5-FU are reportedly factors influencing the flow velocity. However, no study has examined the influence of the site of port insertion. In this study, we investigated factors associated with the flow velocity of 5-FU continuously and intravenously injected by using an infuser in patients with colorectal cancer, among whom the site of port insertion differed. The results showed that the site of port insertion influenced the flow velocity during the continuous intravenous injection of 5-FU using an infuser. It is important to reduce the error in the flow velocity by considering the site of port insertion in addition to the concentration of 5-FU and the air temperature.
要旨
大腸がんに対する化学療法であるフルオロウラシル(5-FU)の46時間持続静注のデバイスとして,加圧式携帯型持続注入器(以下,インフューザー)が用いられるが,流速に誤差が生じることが知られている。投与量または投与終了時間の誤差は5-FUの有効性や安全性に影響する可能性があるため,流速の誤差を極力低減することが求められる。流速への影響因子として希釈液の種類,気温および5-FU濃度などが報告されているが,ポート留置部位の影響について検討した報告はない。そこで,本研究ではポート留置部位が異なる大腸がん患者を対象とし,インフューザーによる5-FU持続静注の流速に影響する因子を検討した。その結果,ポート留置部位はインフューザーによる5-FU持続静注の流速に影響する因子であることが明らかとなった。5-FU濃度や気温以外にポート留置部位も考慮し,流速の誤差を低減することが重要である。